EasyManuals Logo

Agilent Technologies 7890A Troubleshooting

Agilent Technologies 7890A
208 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
Chromatographic Symptoms 3
Agilent 7890 Series Troubleshooting 45
Verify the gas purity. Perform replicate runs, several with a
short interval between them, then several with a large
interval. If the contamination peaks are larger for the runs
made with the longer interval, suspect contaminated gas: the
contamination has more time to deposit onto the column and
liner.
Check for supply tubing and fitting contamination.
If you suspect that there is contamination in the inlet,
perform the bakeout procedure (SS, MMI, PP, COC, PTV, VI).
Verify the operation of the inlets. Clean the inlet and replace
the contaminated inlet parts.
To Clean the Split/Splitless Inlet
To Clean the Purged Packed Inlet
To Clean the Cool On-Column Inlet
To Clean the Septumless Head on the PTV Inlet
To Clean the Multimode Inlet
To Clean the Septum Seat in the Septum Head Assembly
of the PTV Inlet
To Clean the VI
Check the solvent level in the ALS wash bottles.
Replace the ALS syringe if necessary (7693A, 7683).
Check the sample injection volume. Verify that the ALS is
injecting enough sample into the inlet. Use the Solvent Vapor
Volume Calculator to determine how much of the sample
should be injected.
Ghost peaks are sometimes caused by contaminated sample
vials. Try new or clean vials to see if ghost peaks disappear.
Some samples change with heat or ultraviolet light. Check
the sample stability.
Column, method
Perform column maintenance: Bake out contaminants,
remove the contaminated length of column near the inlet,
and reverse and bake out the column as needed.
If you suspect that there is contamination in the column,
perform the bakeout procedure.
Verify that the oven program temperature and time are
sufficient for the samples being injected. Ghost peaks that
are broader than adjacent sample peaks could be from a
previous run.

Table of Contents

Other manuals for Agilent Technologies 7890A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890A and is the answer not in the manual?

Agilent Technologies 7890A Specifications

General IconGeneral
BrandAgilent Technologies
Model7890A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals