46 Agilent 7890 Series Troubleshooting
3 Chromatographic Symptoms
Inspect the column for contamination. High molecular weight
samples that contain residues may cause the syringe, the
inlet liner, or the first few inches of column to become
contaminated.
Install an Agilent column backflush system.
Detector, detector gas supply
Check all gas trap indicators and dates.
Verify the gas purity. Perform replicate runs, several with a
short interval between them, then several with a large
interval. If the contamination peaks are larger for the runs
made with the longer interval, suspect contaminated gas: the
contamination has more time to deposit onto the column and
liner.
Check for supply tubing and fitting contamination.
If you suspect that there is contamination in the detector,
perform the bakeout procedure (FID, TCD, uECD).
Verify the operation of the detectors. Replace the
contaminated detector parts.
To Maintain the NPD Collector, Ceramic Insulators, and
Jet
Maintaining the FID

Table of Contents

Other manuals for Agilent Technologies 7890A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7890A and is the answer not in the manual?

Agilent Technologies 7890A Specifications

General IconGeneral
BrandAgilent Technologies
Model7890A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals