1260 Infinity Binary LC - System User Guide 115
Appendix
6
Setting Up a Method using Edit Entire Method
Autosampler tab
Figure 25 Setup Method screen – High Performance Autosampler tab
•Injection Mode
Injection volume sets the volume to be injected (example 1 µl),
Standard injection indicates that no external needle wash is done,
Injection with needle wash is used to reduce potential carryover. This is
the recommended option and is configured in the next entry.
Needle Wash – if selected above.
Mode determines how the outside of the needle will be rinsed: either
actively in the Flush Port or by dipping into a specified Wash Vial.
Time in seconds the peristaltic pump connected to the flush port will
pump the rinse solvent. It then pumps for a further 15 s to clear the
flush port.
Location determines which vial or well plate will be used if Wash Vial
was selected.
NOTE
Vials should not have a septum i.e., they should be open to avoid transfer of carryover
material on the septum.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 1260 Infinity and is the answer not in the manual?

Agilent Technologies Agilent 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals