EasyManua.ls Logo

Agilent Technologies Agilent 1260 Infinity - Delay Volumes in the Agilent 1260 Infinity Binary LC

Agilent Technologies Agilent 1260 Infinity
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42 1260 Infinity Binary LC - System User Guide
3 Optimization of the Agilent 1260 Infinity Binary LC
How to Configure the Optimum Delay Volume
Delay Volumes in the Agilent 1260 Infinity Binary LC
Table 4 on page 42 and Table 5 on page 42show the component volumes
which contribute to system delay volume in the Agilent 1260 Infinity
Binary LC System .
Switching between configurations can be done in two ways:
manually, by disconnecting and reconnecting capillaries
automatically, using a 600 bar 2PS/6PT valve (optional)
Tabl e 4 Delay volumes of the 1260 Infinity Binary LC modules
Components Delay Volume (µL)
Binary Pump
1
1
in low delay volume configuration with bypassed damper and mixer
120
Binary Pump
2
2
in standard delay volume configuration
600 – 800
Low volume mixer 200
Mixer 400
Autosampler 270
Low dispersion heat exchanger 1.6
Built-in heat exchanger 3 and 6
Tabl e 5 Delay volumes of 1260 Infinity Binary LC configurations
System Configuration Delay Volume (µL)
Low delay volume configuration Pump: 120
Autosampler:
Medium delay volume configuration Pump: 320
Standard delay volume configuration Pump: 600 – 800

Table of Contents

Related product manuals