56 1260 Infinity Binary LC - System User Guide
3 Optimization of the Agilent 1260 Infinity Binary LC
How to Achieve Higher Resolution
Optimum Instrument Configuration for High Resolution
Column Compartment
The column compartment can be used in its standard version for 4.6 mm
inner diameter columns. At flow rates above 2 mL/min, and temperatures
above 60 °C, the column effluent should be cooled down to the
temperature of the detector by using the 1.5 µL cooler/heater in the
column compartment. This ensures lowest noise level with UV detectors,
even at 5 mL/min and 80 °C (see Figure 18 on page 56 ).
Figure 18 Influence of post column cooling (PCC) on baseline noise
If 2.1 mm inner diameter columns are used at low flow rates, the small
heating device should be used for lowest extra- column volume.
L^i]edhiXdajbcXdda^c\
H$C[dgeZV`)2)-'
CdedhiXdajbcXdda^c\
H$C[dgeZV`)2(.'
Cdgb#
b^c
Cd^hZEidE2&#*.%*b6J
Cd^hZEidE2%#-'-.b6J
8dda^c\cZZYZY3'ba$b^c

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 1260 Infinity and is the answer not in the manual?

Agilent Technologies Agilent 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals