94 1260 Infinity Binary LC - System User Guide
5 Quick Start Guide
Data Acquisition in Method and Run Control View
Running the Method for a Single Injection
This section shows how to run a single injection of the isocratic checkout
sample using the conditions entered in the previous section.
Analyses on the ChemStation can be run in two modes:
•Run Method — single injections, for example, in interactive method
development, using the parameter settings current at the time;
•Run Sequence automated series of injections from multiple vials,
possibly with multiple methods. For further details refer to the
ChemStation manuals.
1 Click the Select Run Method Task icon .
2 If required method conditions are not currently loaded, select Method >
Load Method or the icon under the menu bar to load them.
3 Place the sample vial in position 1 — this is the front position of the 10
x 2 ml vial positions on the right hand side of the sample tray.
4 Select Run Control > Sample Info, and enter the required sample
parameters.
For example:
Subdirectory (optional)
Name Pattern
Vial/Location
Sample name
Comment
5 If the system is already equilibrated and the baseline is stable, click Run
Method in the Sample Info page to start the injection. Alternatively, click
OK, and when ready, select Run Control > Run Method. The item will be
added to the Run Queue and then automatically started from there.
NOTE
If changes have been made to a loaded method and not yet saved, this is indicated by a
yellow asterisk on the method status icon. The injection can be made without first saving
those parameters changes. The ChemStation always stores a copy of the acquisition
parameters in the data file as ACQ.TXT to ensure the preservation of the original method
parameters.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 1260 Infinity and is the answer not in the manual?

Agilent Technologies Agilent 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals