136 1260 Infinity Binary LC - System User Guide
6 Appendix
Setting Up a Method using Edit Entire Method
Run Time Checklist
The Run Time Checklist can also be directly accessed through the menu
Method > Run Time Checklist... or by clicking on the Run Time Checklist icon at
the top right of the screen.
Figure 35 Run Time Checklist Screen
The Run Time Checklist selects whether the method should run both data
acquisition and data analysis and also offers an opportunity to link macro
commands or programs into the work flow at various points. In most
cases the Data Acquisition and Standard Data Analysis check- boxes are
checked. If no data analysis is required, for example, in a series of method
development runs, Standard Data Analysis can be unchecked so that no
report is produced and the data can be visually assessed later in the Data
Analysis view.
To link in a macro program at one of the access points the relevant box is
checked and the name of the macro is typed in the text box to the right.
The software looks for the macro in the C:\Chem32\Core directory; include
the path if it is located elsewhere.