Optimization of the Agilent 1260 Infinity Binary LC
3
How to Achieve Higher Sensitivity
Choosing a Flow Cell
Several flavors of the Max- Light Cartridge Flow Cell are available, see
(Table 7 on page 63).
Normal Applications
The Max- Light Cartridge Cell ( 10 mm, V(s) 1.0 µL) (G4212- 60008) covers
a wide range of applications:
all column diameter down to at least 2.1 mm ID or even less
applications with peak dispersion (Peakwidth x flow) down to ~L
[example: pw = 0.04 min at flow = 0.1 mL/min gives peak dispersion of
0.04 min x 0.1 mL/min = 0.004 mL = 4 µL]
High Sensitiviy
If higher sensitivity is necessary, the Max- Light Cartridge Cell ( 60 mm,
V(s) 4.0 µL) (G4212- 60007) can be used. This cell enhances the detector
by lowering the limit of detection (LOD) by a factor of about 3 (depending
on the application).
Tabl e 7 Specifications for Max-Light Cartridge Flow Cells
Cartridge Cells Max-Light Cartridge Cell ( 10 mm, V(s) 1.0 µL) (G4212-60008)
Max-Light Cartridge Cell Bio-inert ( 10 mm, V(s) 1.0 µL) (G5615-60018)
Max-Light Cartridge Cell ( 60 mm, V(s) 4.0 µL) (G4212-60007)
Max-Light Cartridge Cell Bio-inert ( 60 mm, V(s) 4.0 µL) (G5615-60017)
HDR Max-Light Cartridge Cell ( 3.7 mm, V(s) 0.4 µL) (G4212-60032)
ULD Max-Light Cartridge Cell ( 10 mm, V(s) 0.6 µL) (G4212-60038)
Max-Light Cartridge Test Cell (G4212-60011)
Maximum
pressure
70 bar (1015 psi) Maximum Operating Pressure (MOP)
1
150 bar (2175 psi) Maximum Incidential Pressure (MIP)
2
1
Maximum Operating Pressure (MOP): The maximum pressure at which the system can operate
continuously under normal conditions.
2
Maximum Incidential Pressure (MIP): The maximum pressure which the system can experience
during a short time.
pH range 1.0-12.5 (solvent dependent)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 1260 Infinity and is the answer not in the manual?

Agilent Technologies Agilent 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals