54 1260 Infinity Binary LC - System User Guide
3 Optimization of the Agilent 1260 Infinity Binary LC
How to Achieve Higher Resolution
was used in the decision on phases, it is likely that short columns were used 
for fast analysis in each step of the scouting. 
The resolution equation shows that the next most significant term is the 
plate count or efficiency, N, which can be optimized in a number of ways. 
N is inversely proportional to the particle size and directly proportional to 
the length of a column and so smaller particle size and a longer column 
will give a higher plate number. The pressure rises with the inverse square 
of the particle size and proportionally with the length of the column. 
Resolution increases with the square root of N so doubling the length of 
the column will increase resolution by a factor of 1.4. What is achievable 
depends on the viscosity of the mobile phase as this relates directly to the 
pressure. Methanol mixtures will generate more back pressure than 
acetonitrile mixtures. Acetonitrile is often preferred because peak shapes 
are better and narrower in addition to the lower viscosity but methanol 
generally yields better selectivity (certainly for small molecules less than 
about 500 Da). The viscosity can be reduced by increasing the temperature 
but it should be remembered that this can change the selectivity of the 
separation. Experiment will show if this leads to increase or decrease in 
selectivity. As flow and pressure are increased, it should be remembered 
that frictional heating inside the column will increase. That can lead to 
slightly increased dispersion and possibly a small selectivity change both 
of which could be seen as a reduction in resolution. The latter case might 
be offset by reducing the temperature of the thermostat by a few degrees 
and again experiment will reveal the answer. 
The van Deemter curve shows that the optimum flow rate through an STM 
column is higher than for larger particles and is fairly flat as the flow rate 
increases. Typical, close to optimum, flow rates for STM columns are: 
2 ml/min for 4.6 mm i.d.; and 0.4 ml/min for 2.1 mm i.d. columns.