1260 Infinity Binary LC - System User Guide 79
System Setup and Installation
4
Installing the Modules
Changing Solvents
1 If the channel is not filled with buffer, proceed to step 4.
2 Place the solvent intake filter into a bottle of water.
3 Flush the channel at a flow rate suitable for the installed tubing
(typically 3 – 5 mL/min) for 10 min.
4 Modify the flow path of your system as required for your application.
5 Replace the solvent bottle by a bottle of isopropanol.
6 Flush the channel at a flow rate suitable for the installed tubing
(typically 3 – 5 mL/min) for 5 min.
7 Swap the bottle of isopropanol with a bottle of solvent for your
application.
8 Repeat steps 1 to 7 for the other channel(s) of the pump.
9 Install the desired column, set the required composition and flow rate
for your application and equilibrate the system for approx. 10 minutes
prior to starting a run.
When When the solvent of a channel is to be replaced by another solvent that is not compatible (solvents
are immiscible or one solvent contains a buffer), it is necessary to follow the procedure below to
prevent clogging of the pump by salt precipitation or residual liquid droplets in parts of the system.
Parts required # p/n Description
1 Purging solvent(s), see Ta b l e 8 on page 80
1 5022-2184 Union ZDV
Preparations Remove the column and replace it by a ZDV fitting
Prepare bottles with appropriate intermediate solvents (see Ta b l e 8 on page 80)
CAUTION
Buffer salt of aqueous buffers may precipitate in residual isopropanol.
Capillaries and filter may be clogged by precipitating salt.
Flush solvent lines containing high concentration of salts first with water before
introducing organic solvent.
Do not perform steps 5 to 7 for channels running with aqueous buffer as solvent.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 1260 Infinity and is the answer not in the manual?

Agilent Technologies Agilent 1260 Infinity Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 1260 Infinity
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals