Return Loss Meters Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 403
error message E10081: Safety comparator <a> test failed (b, c)
description Checks whether laser current safety comparator includes at the
right conditions a safety interrupt at the CPU
test value a laser number
b shows whether the interrupt flag was raised when the
current was lower than the limit
c shows whether the interrupt flag was raised when the
current was higher than the limit
action Please return the unit to your local service center for repair