Optical Switch modules Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 419
error message KeyDisable(): ERROR
KeyDisable(): disabledCnt >= disabledMax
KeyDisableTask(): received wrong channel number.
KeyDisableTask(): disabledCnt <0.
KeyDisableTask(): couldn’t send myself a msg.
KeyEnablePortF(): wrong interrupt number
KeyDisablePortF(): wrong interrupt number
description Problem when enabling or disabling a front panel key. Probably a
calibration data or firmware problem.
test value
action Please return the unit to your local service center for repair
error message KeySelftest() called without calling KeyInit() before
KeySelftest(): KeyDisable() failed
KeySelftest(): selftestPre() failed
KeySelftest(): key pressed during selftest or hardware problem
KeySelftest(): selftestPost() failed
KeySelftest(): KeyEnable() failed
KeySelftestExecPortF(): wrong interrupt number
KeySelftestPreTPU(): FreeTpultc() failed
KeySelftestExecTPU(): GetTpuDioState() failed
KeySelftestPostTPU(): AllocTpultc() failed
Selftest (stAll, 0) failed
description Selftest problem. Possible causes:
- hardware failure
- key pressed during selftest
- calibration data or firmware incorrect
test value
action Please return the unit to your local service center for repair