Optical Switch modules Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 419
error message KeyDisable(): ERROR
KeyDisable(): disabledCnt >= disabledMax
KeyDisableTask(): received wrong channel number.
KeyDisableTask(): disabledCnt <0.
KeyDisableTask(): couldn’t send myself a msg.
KeyEnablePortF(): wrong interrupt number
KeyDisablePortF(): wrong interrupt number
description Problem when enabling or disabling a front panel key. Probably a
calibration data or firmware problem.
test value
action Please return the unit to your local service center for repair
error message KeySelftest() called without calling KeyInit() before
KeySelftest(): KeyDisable() failed
KeySelftest(): selftestPre() failed
KeySelftest(): key pressed during selftest or hardware problem
KeySelftest(): selftestPost() failed
KeySelftest(): KeyEnable() failed
KeySelftestExecPortF(): wrong interrupt number
KeySelftestPreTPU(): FreeTpultc() failed
KeySelftestExecTPU(): GetTpuDioState() failed
KeySelftestPostTPU(): AllocTpultc() failed
Selftest (stAll, 0) failed
description Selftest problem. Possible causes:
- hardware failure
- key pressed during selftest
- calibration data or firmware incorrect
test value
action Please return the unit to your local service center for repair

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 8163A and is the answer not in the manual?

Agilent Technologies Agilent 8163A Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 8163A
CategoryMultimeter
LanguageEnglish

Related product manuals