Specifications
16 Agilent Cary 60 Spectrophotometer User’s Guide
Environmental Conditions
No draughts No corrosive fumes No vibration
NOTE
For optimum analytical performance, it is recommended that the ambient
temperature of the laboratory be between 20 and 25 °C and be held constant to
within
±2 °C throughout the entire working day.
The area should have a dust-free, low humidity atmosphere. Air
conditioning is recommended. The room should be temperature-
controlled if your analyses are particularly sensitive.
Condition Altitude (m, ft) Temp. (°C, °F) Humidity (%RH)
non-condensing
Non-operating (transport) 0–4600, 0–15000 -40–70, -40–158 <90
Operating within performance
specifications
0–3100, 0–10000 5–40, 41–104 15–80