Troubleshooting
Agilent Cary 60 Spectrophotometer User’s Guide 41
Wavelength Calibration
Problem
The position of an absorption (transmission or reflectance) peak
appears to shift over a prolonged period of time.
Solution
Restart the Cary 60. The calibration procedure occurs automatically
on start up. Agilent recommends that as part of your standard
operating procedure you restart the Cary 60 monthly. Perform the
Instrument Performance Test suite in ‘Validate’ to ensure the
instrument results meet specification.
Front Panel LED Troubleshooting
LED Won’t Light Up
No power to the instrument.
Check power cables and power point switches (if applicable).
Solid Red LED
The instrument is available to use but the previous initialization or
calibration has failed.
Initialization or calibration failed.
Check or perform the following:
The sample compartment for items blocking the beam.
The cuvette is inserted correctly and clean.
Reboot the Cary 60.
Contact your local Agilent field service engineer or
representative.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 60 and is the answer not in the manual?

Agilent Technologies Cary 60 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 60
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals