Maintenance and Spare Parts
Agilent Cary 60 Spectrophotometer User’s Guide 43
6. Maintenance and Spare Parts
Maintenance 43
Spare Parts 44
Maintenance
Any procedures not specifically mentioned in this chapter or in the
Cary WinUV Help should be carried out only by Agilent field service
engineers.
WARNING
Eye Hazard
This instrument contains an intense light source. Direct viewing of the light
source will cause eye damage. Operators and other unauthorized personnel
must NEVER remove the main cover.
NOTE
This section refers only to maintenance procedures for the instrument. You
should refer to your PC and printer manuals for their maintenance procedures,
and to the Cary WinUV Help for the maintenance procedures for any accessories
you ordered.
Any spills in the sample compartment should be wiped up
immediately.
The exterior surfaces of the Cary 60 spectrophotometer should be
kept clean. All cleaning should be done with a soft cloth. If necessary,
this cloth can be dampened with water or a mild detergent. Do not
use organic solvents or abrasive cleaning agents.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 60 and is the answer not in the manual?

Agilent Technologies Cary 60 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 60
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals