Troubleshooting
Agilent Cary 60 Spectrophotometer User’s Guide 35
5. Troubleshooting
Access Denied 35
No Green Light Indicating that the Cary 60 is Powered 37
Start Button Replaced with Connect Button 37
Absorbance is 10 Abs and Fluctuates Wildly During a Scan
39
Instrument Performance Testing 40
Wavelength Calibration 41
Front Panel LED Troubleshooting 41
This chapter contains troubleshooting information to help you solve
various problems you may encounter when setting up or using your
Cary 60 hardware. If the suggested solutions do not fix the problem,
please contact your local Agilent field service engineer or
representative.
Access Denied
Problem
During installation of the Cary WinUV software, you see an ‘Access is
denied’ message.
Solution
You must be logged on with administrator rights to install the Cary
WinUV software.
1 Click the Windows Start button > Log Off.
2 Log on as an Administrator or ask your Administrator to log on
for you.