Specifications
18 Agilent Cary 60 Spectrophotometer User’s Guide
Electrical Specifications
Mains Supply
A standard 3.2 A/12 V plug pack is provided. Power cords are
provided based on the user’s country requirements. Only the
supplied power supply is to be used with this equipment.
Table 1. System electrical specifications
System unit Required supply voltage
Spectrophotometer
90-265 V AC
Frequency 47-63 Hz
The maximum power consumption is 60 VA, 38 W, 130 BTU/hr.
The installation of electrical power supplies must comply with the
rules and/or regulations imposed by local authorities responsible for
the supply of electrical energy to the workplace.
WARNING
Shock Hazard
Danger of electrocution. Good electrical grounding is essential to avoid
potentially serious shock hazards. A 3-wire outlet with ground connection
must be provided for the Cary 60. Make certain that power outlets are earth-
grounded at the grounding pin.
All power supplies for the Cary 60 must be single-phase, AC voltage,
three-wire system (active, neutral, earth) and should be terminated
at an appropriate power outlet receptacle that is within reach of the
power cord. For safety reasons, a separate power outlet receptacle
should be provided for each unit in the system. The use of extension
cords or outlet adaptors is not recommended.