Chapter 5 Error Messages
Execution Errors
120
-158
String data not allowed
A character string was received but is not allowed for the command. Check the list
of parameters to verify that you have used a valid parameter type.
Example: TRIG:DEL ’zero’
-160 to -168
Block data errors
The power supply does not accept block data.
-170 to -178
Expression errors
The power supply does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (GET) or *TRG was received but the trigger was ignored.
Make sure that the trigger source should be selected to the bus and the trigger
subsystem should be initiated by INIT[:IMM] command.
-213
[1]
Trigger ignored
An INITiate command was received but could not be executed because a
measurement was already in progress. Send a device clear to halt a measurement in
progress and place the power supply in the “idle” state.
-221
Init ignored
Indicates that a legal program data element was parsed but could not be executed due
to the current device state.
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example: TRIG:DEL -3
-223
Too much data
A character string was received but could not be executed because the string length
was more than 40 characters. This error can be generated by the
CALibration:STRing command.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for the command.
You may have used an invalid parameter choice.
Example: DISP:STAT XYZ (XYZ is not a valid choice).
[1]This error message is only applicable for serial MY53xx6xxx.