122 Chapter 2
Instrument Functions: A - L
File
Instrument Functions: A - L
NOTE You are always safe pressing Save Now without entering a file name, because the
auto-generated file name never conflicts with an existing file.
If the Path: field above the directory box is empty when pressing Save Now,
the status line will display the error message: Unable to save file,
invalid path. In this case, please select a drive.
Key Path:
File, Save
Remote Command:
:MMEMory:STORe:SCReen <‘file_name’>
:MMEMory:STORe:STATe 1,<‘file_name’>
The MMEM:STOR:STAT command only saves the state of the Signal Analysis mode. Use *SAV
to save all mode states.
:MMEMory:STORe:TRACe <label>,<‘file_name’>
This command is not available when you are outside the Spectrum Analysis mode and
working in optional measurement modes.
:MMEMory:STORe:LIMit LLINE1|LLINE2,<‘file_name’>
:MMEMory:STORe:CORRection ANTenna|CABLe|OTHer|USER,<‘file_name’>
:MMEMory:STORe:RESults <‘file_name’>
*SAV <register#>
Remote Command Notes: For the MMEM:STOR:TRAC <label>,<“file_name”> command:
Trace labels are: TRACE1|TRACE2|TRACE3|ALL
The file name must have a file extension of .TRC or .CSV. The file
extension determines whether a trace is stored (.CSV), or a trace with
its state (.TRC), are stored.
The <‘file_name’> must include the complete path, for example
‘C:\MYTRACE.TRC’. Lowercase characters are interpreted as uppercase.
These commands will fail if the <‘file_name’> already exists.
For the MMEM:STOR:LIM LLINE2,”C:mylimit.lim” command, there is no
short form for parameters LLINE1|LLINE2.