Chapter 2 149
Instrument Functions: A - L
FREQUENCY / Channel
Instrument Functions: A - L
2.7.6 Signal Track
When a marker is placed on a signal and Signal Track is pressed, the marker will remain on
the signal while the analyzer retunes the center frequency to the marker frequency. The
analyzer will keep the signal at the center of the display, as long as the amplitude of the
signal does not change by more than 3 dB from one sweep to another. If no marker is
active, pressing
Signal Track to On will activate a marker, perform a peak search, and
center the marker on the display.
If the signal is lost, an attempt will be made to find it again and continue tracking. If there
are other signals on screen near the same amplitude, one of them may be found instead.
Signals near 0 Hz cannot be tracked effectively as they cannot be distinguished from the
LO feedthrough, which is excluded by intent from the search algorithm.
When
Signal Track is On and the span is reduced, an automatic zoom is performed and the
span is reduced in steps so that the signal remains at the center of the display. If the span
is zero, signal track cannot be activated.
NOTE This function is intended to track signals with a frequency that is changing
(drifting), and an amplitude that is not changing. It keeps tracking if in
continuous-sweep mode. If in single-sweep mode, the analyzer only does one
center frequency adjustment as necessary.
Signal tracking can also be used with the
(delta) pair marker function. This could be
used to measure the amplitude at a frequency offset, relative to the fundamental signal
amplitude, even if the fundamental is drifting. In this situation, the tracking function
will be done on the delta marker, not on the reference marker. So you would want to put
the ref marker on the frequency+offset, and put the
marker on the fundamental signal
peak. Then turn on signal tracking. The frequency difference between the two markers will
stay fixed. The
marker, on the fundamental, will track to the center of the display. The
marker readout value will show the amplitude delta between the two markers.

Table of Contents

Other manuals for Agilent Technologies E4448A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4448A and is the answer not in the manual?

Agilent Technologies E4448A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4448A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals