Chapter 6 375
Using the STATus System
STATus Subsystem
Using the STATus System
Key Type: There is no equivalent front-panel key.
Factory Preset: 32767 (all 1s)
Range: 0 to 32767
Questionable Calibration Register
Questionable Calibration Condition
:STATus:QUEStionable:CALibration:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Calibration Condition register.
NOTE The data in this register is continuously updated and reflects the
current conditions.
Key Type: There is no equivalent front-panel key.
Questionable Calibration Enable
:STATus:QUEStionable:CALibration:ENABle <number>
:STATus:QUEStionable:CALibration:ENABle?
This command determines which bits in the Questionable Calibration
Condition Register will set bits in the Questionable Calibration Event
register, which also sets the Calibration Summary bit (bit 8) in the
Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Key Type: There is no equivalent front-panel key.
Example STAT:QUES:CAL:ENABLE 16384 could be used if you
have turned off the automatic alignment and you want
to query if an alignment is needed.
Factory Preset: 32767 (all 1s)
Range: 0 to 32767

Table of Contents

Other manuals for Agilent Technologies E4448A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4448A and is the answer not in the manual?

Agilent Technologies E4448A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4448A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals