Chapter 4 283
Instrument Functions: P - Z
Trig
Instrument Functions: P - Z
4.11.3 Line
Sets the trigger to start a new sweep/measurement to be synchronized with the next cycle
of the line voltage.
Key Path:
Trig
Couplings/
Dependencies: Line trigger is not available when operating from a dc power source.
Remote Command:
See Trig on page 281
Example: TRIG:SOUR LINE selects line triggering.
4.11.4 Ext Front (Ext Trig In)
Sets the trigger to start a new sweep/measurement whenever the external voltage
(connected to EXT TRIGGER INPUT on the front panel) passes through approximately 1.5
volts.
Key Path:
Trig
State Saved: Saved in instrument state
Factory Preset: 1.5 V
Remote Command:
See
Trig on page 281
Example: TRIG:SOUR EXT to select front panel external triggering.
4.11.5 Ext Rear (Trigger In)
Sets the trigger to start a new sweep/measurement whenever the external voltage
(connected to TRIGGER IN on the rear panel) passes through approximately 1.5 volts.
Key Path:
Trig
State Saved: Saved in instrument state
Factory Preset: 1.5 V
Remote Command:
See
Trig on page 281
Example: TRIG:SOUR EXT2 to select rear panel external triggering.

Table of Contents

Other manuals for Agilent Technologies E4448A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4448A and is the answer not in the manual?

Agilent Technologies E4448A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4448A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals