EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E4448A

Agilent Technologies E4448A User Manual

Agilent Technologies E4448A
434 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #366 background imageLoading...
Page #366 background image
366 Chapter 6
Using the STATus System
Common Using the STATus System IEEE Commands
Using the STATus System
For example, if you want to verify the completion of both calibration
and waiting for trigger set :STAT:OPER:ENAB 33 and monitor any
changes.
Key Type: There is no equivalent front-panel key.
Query Instrument Options
*OPT?
For ESA and PSA analyzers: This command is not implemented.
However the information is available remotely by using SCPI
commands. See :SYSTem:OPTions? in the SYSTem subsystem.
Power-On Status Clear
*PSC
*PSC?
Sets the state of the power-on status clear flag. This command allows
you to specify if the service request enable register and the event status
enable register should be cleared at power up.
The query returns the contents of the power-on status clear flag.
Key Type: There is no equivalent front-panel key.