100 1100 Series WS MWS Reference Manual
3 Troubleshooting and Test Functions
4. Rheodyne valve error
Turn the system off and on twice
Check the connectors on the SUD board
Check the sampling unit connector on the main board
Check the connector on the sampling unit
Change the rheodyne valve (0101-0921)
Change the sampling unit (G1367-60008)
Change the main board (G1367-69500)
5. Needle lock error
Check the connectors on the SUD board
Check the sampling unit connector on the main board
Check the connector on the sampling unit
Change the sampling unit (G1367-60008)
Change the main board (G1367-69500)