EasyManua.ls Logo

Agilent Technologies G1367A

Agilent Technologies G1367A
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
54 1100 Series WS MWS Reference Manual
2 Optimizing Performance
Choice of Seat Capillary
Different models of seat capillaries are available for the well-plate sampler and
the micro well-plate sampler:
For the Well-plate
Sampler
The needle seat assembly includes the needle seat and the seat capillary. The
part number for this assembly is: G1367-87101.
For the Micro
Well-plate
Sampler
The needle seat assembly is made up of two parts:
needle seat: G1377-87101.
seat capillary
choice of: G1375-87317 (100 µm 150 mm)
G1375-87300 (50 µm 150 mm)
G1375-87317 (100 µm) is the capillary preinstalled in micro well-plate and
thermostatted micro well-plate samplers upon delivery.
This capillary is recommended for applications with a 0.3 mm column or
higher. It provides less plugging of the capillary in general and especially with
biological samples. For small K’ this capillary can provide a higher peak width
for isocratic analysis.
G1375-87300 (50 µm) is available as a spare part and is recommended for
applications with a 0.3 mm column or smaller. This capillary gives full
chromatographic performance.

Table of Contents

Related product manuals