72 1100 Series WS MWS Reference Manual
3 Troubleshooting and Test Functions
Front Door Error
The front door and/or the SLS board are damaged.
Probable Causes
The sensor on the SLS board is defective.
The door is bent or the magnet is misplaced/broken.
Suggested Actions
Exchange the door.
Exchange the SLS board.
Exchange the MTP board.