Troubleshooting and Test Functions 3
1100 Series WS MWS Reference Manual 79
Needle Carrier Failed
The needle carrier on the Sample Transport Assembly failed to move correctly.
Probable Causes
Defective Z-motor.
Vial pusher blocked.
Bad needle carrier positioning in X or Theta.
Defective vial pusher sensor.
Defective MTP main board.
Possible Actions
Exchange the needle carrier assembly.
Exchange the sample transport assembly.
Exchange the MTP main board.