Troubleshooting and Test Functions 3
1100 Series WS MWS Reference Manual 85
Peristaltic Pump Error
The peristaltic pump motor in the well-plate sampler has failed.
The current on the motor is used by the MTP board to monitor the speed of the
peristaltic pump motor. If the current falls below a certain value, the error
message is generated.
Probable Causes
Defective motor.
Defective SUD board.
Defective MTP main board.
Suggested actions
Exchange the peristaltic pump motor.
Exchange the SUD board.
Exchange the MTP main board.