Repairing the Sampler 4
1100 Series WS MWS Reference Manual 121
9 Install this stator head/face assy on the injection valve. Tighten the screws
alternately with the 9/64 inch wrench until the stator head is secure.
10 Reconnect all the capillaries to the injection valve ports with the 1/4 inch
wrench.
11 Close the front cover.