Repairing the Sampler 4
1100 Series WS MWS Reference Manual 125
10 Pinch the holder clamp, pull back and remove the needle assembly with the
loop capillary from the needle carrier.
11 Attach the 5/16 inch wrench to hold position at the needle assembly. Use
the 4 mm wrench to loosen the fitting of the loop capillary.
12 Pull the loop capillary out from the needle assembly.