Repairing the Sampler 4
1100 Series WS MWS Reference Manual 127
10 Remove the silicon safety tube from the needle.
11 Replace the plate tray in the tray base. Re-install the side door and close the
front door.
12 In the user interface close the “Change Loop Capillary" function and exit
themaintenance mode.