150 1100 Series WS MWS Reference Manual
4 Repairing the Sampler
Installing the SUD board
1 Install the new board. Ensure the board is positioned between the board
guide.
2 Secure the board in place with the two screws.
3 Reconnect all the connectors to the new board.
4 Install the sampling unit, (see “Installing the sampling unit" on page 135).
5 Install the sample transport, (see “Installing the sample transport" on
page 133)
6 Install the top cover, top plate, and foam (see “Top Cover and Foam" on
page 132).
7 Replace the plate tray in the tray base.
8 Turn ON the sampler, close the front door.
9 The system will reset automatically.
10 In the user interface perform a “Sample Transport Self Alignment”.