Repairing the Sampler 4
1100 Series WS MWS Reference Manual 153
Installing the power supply
1 Insert the coupler onto the switch of the new power supply.
2 Install the power supply in the module.
3 Connect the light pipe to the coupler.
4 Install the bottom foam. Ensure the cables are positioned properly.
5 Install the exhaust fan, (see “Installing the exhaust fan" on page 145).
6 Install the MTP board, (see “Installing the MTP main board" on page 147).
7 Install the main fan, (see “Installing the main fan" on page 144).
8 Install the sampling unit, (see “Installing the sampling unit" on page 135).
9 Install the sample transport, (see “Installing the sample transport" on
page 133)
10 Install the top cover, top plate, and foam (see “Top Cover and Foam" on
page 132).
11 Replace the plate tray in the tray base.
12 Turn ON the sampler, close the front door.
13 The system will reset automatically.
14 In the user interface perform a “Sample Transport Self Alignment”.