SCHEDULED MAINTENANCE: Maintaining the nominal state of operation.
Table 7
Oil level checking
Daily (before every starting)
Oil change
8.000 hours (light applications)
Anti-Suckback valve checking
6 months
Fan cover cleaning
6 months
UNSCHEDULED MAINTENANCE: Restoring the nominal state of operation.
Exhaust filter replacement if oil mist at exhaust or yearly.
NOTE
The frequency with which repairs are performed depends on the process and
presence of substances that shorten pump life (dust, abrasives, solvents, water,
chemically aggressive substances).
Use only AVF Platinum oil. The use of oils other than the one foreseen alters the
pump performance, could damage the pump and in any case invalidate the
warranty.