EasyManua.ls Logo

Agilent Technologies N5242A - Page 58

Agilent Technologies N5242A
342 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-26 Service Guide N5242-90001
Tests and Adjustments PNA Series Microwave Network Analyzers
System Verification N5242A
Interpreting the Verification Results
Figure 3-11 shows an example of typical verification results with Print Tabular Data
selected in the Printer Output area of the System Verification dialog box.
At the top of the printed output is the name of the device, the serial number of the device,
and the date tested.
Each S-parameter measurement result is printed with frequency tested, lower and upper
limit lines, the measured data, and the result of the test.
Figure 3-11 Example of Printed Tabular Verification Results

Table of Contents