Service Guide N5242-90001 3-27
PNA Series Microwave Network Analyzers Tests and Adjustments
N5242A System Verification
Figure 3-12 shows an example of typical verification results with Print Graphs selected in
the Printer Output area of the System Verification dialog box. The printed graphical
results show the following:
the name of the device measured
the serial number of the device
the parameters measured
Results of the measurements. Labeled as A in Figure 3-12.
Data measured at the factory from the verification kit. Labeled as B in Figure 3-12.
Upper and lower limit points as defined by the total system uncertainty system.
Labeled as C in Figure 3-12.
Figure 3-12 Example of Printed Graphical Verification Results