6 Language Dictionary
78 Series N6700 User’s Guide
SCPI Command Description
[SOURce:] (continued)
VOLTage
[:LEVel]
[:IMMediate][:AMPLitude] <NRf+>, (@chanlist) Sets the output voltage
:TRIGgered [:AMPLitude] <NRf+>, (@chanlist) Sets the triggered output voltage
:MODE FIXed | STEP | LIST, (@chanlist) Sets the voltage trigger mode
:PROTection
[:LEVel] <NRf+>, (@chanlist) Sets the over-voltage protection level
:RANGe <NRf+>, (@chanlist) Sets the output voltage range (Agilent N6761A/62A)
:SLEW <NRf+> | INFinity, (@chanlist) Sets the output voltage slew rate
STATus
:OPERation
[:EVENt]? (@chanlist) Returns the value of the operation event register
:CONDition? (@chanlist) Returns the value of the operation condition register
:ENABle <NRf>, (@chanlist) Enables specific bits in the Event register
:NTRansition <NRf>, (@chanlist) Sets the Negative transition filter
:PTRansition <NRf>, (@chanlist) Sets the Positive transition filter
:PRESet Presets all enable and transition registers to power-on
:QUEStionable
[:EVENt]? (@chanlist) Returns the value of the questionable event register
:CONDition? (@chanlist) Returns the value of the questionable condition register
:ENABle <NRf>, (@chanlist) Enables specific bits in the Event register
:NTRansition <NRf>, (@chanlist) Sets the Negative transition filter
:PTRansition <NRf>, (@chanlist) Sets the Positive transition filter
SYSTem
:CHANnel
[:COUNt]? Returns the number of output channels in a mainframe
:MODel? (@chanlist) Returns the model number of the selected channel
:OPTion? (@chanlist) Returns the option installed in the selected channel
:SERial? (@chanlist) Returns the serial number of the selected channel
:COMMunicate
:RLSTate LOCal | REMote | RWLock Specifies the Remote/Local state of the instrument
:TCPip:CONTrol? Returns the control connection port number
:ERRor? Returns the error number and error string
:GROup
:CATalog? Returns the groups that have been defined
:DEFine (@chanlist) Group multiple channels together to create a single output
:DELete (channel) Removes the specified channel from a group
:ALL Ungroups all channels
:PASSword:FPANel:RESet Resets the front panel lock password to zero
:REBoot Returns the unit to its power-on state
:VERSion? Returns the SCPI version number
TRIGger
:ACQuire (Acquire commands only on Agilent N6761A/62A and Option 054)
[:IMMediate] (@chanlist) Triggers the measurement immediately
:SOURce BUS | PIN<pin> | TRAN<chan>, (@chanlist) Sets the measurement trigger source
:TRANsient
[:IMMediate] (@chanlist) Triggers the output immediately
:SOURce BUS | PIN<pin> | TRAN<chan>, (@chanlist) Sets the output trigger source