Specifications and Characteristics 6
U3402A User’s and Service Guide 103
Audible tone
Continuous beep for continuity and single tone for normal forward-biased diode or
semiconductor junction
Input protection
500 V RMS on all ranges
Resistance/continuity (2-wire W)
Measurement method
2-wire Ohms
Test current
Approximately 0.5 mADC
Open-circuit voltage
Limited to < +5 VDC
Audible tone
Continuous beep for continuity and single tone for normal forward-biased diode or
semiconductor junction
Zeroing error
0.05 W or less (excluding test lead resistances) in each range when Rel operation is used
Input protection
500 V RMS on all ranges
Frequency
Measurement method
Reciprocal counting technique. AC coupled input using AC voltage function
Crest factor
Maximum 3:0 at full scale
Signal level
10% of range to full scale input on all ranges
Auto or manual range selection
Table 6 - 9 Supplemental measurement specifications (continued)

Table of Contents

Other manuals for Agilent Technologies U3402A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U3402A and is the answer not in the manual?

Agilent Technologies U3402A Specifications

General IconGeneral
BrandAgilent Technologies
ModelU3402A
CategoryMultimeter
LanguageEnglish

Related product manuals