EasyManua.ls Logo

Agilent Technologies U3402A

Agilent Technologies U3402A
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Getting Started 1
U3402A User’s and Service Guide 15
Table 1 - 3 Input terminal for different measurement functions
Measurement function Input terminal Overload protection
DC voltage (VDC)
V W Hz (Hi)
Lo
1000 VDC
AC voltage (VAC), frequency (Hz)
750 VAC RMS, 1100 V PEAK,
2x10
7
V-Hz normal mode, or
1x10
6
V-Hz common mode
Miliampere (mA), frequency (Hz) mA 1200 mADC or AC RMS
12A, frequency (Hz) 12A
12 ADC or AC RMS for
continuous 30 seconds, or
10 ADC or AC RMS
2-wire resistance (W (2W))
V W Hz 500 VDC or AC RMS
Diode test, continuity test
4-wire resistance (W (4W)) Hi 250 VDC or AC RMS
All functions Any terminal to earth 1000 VDC or AC PEAK

Table of Contents

Other manuals for Agilent Technologies U3402A

Related product manuals