EasyManua.ls Logo

Agilent Technologies U3402A

Agilent Technologies U3402A
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1Getting Started
28 U3402A User’s and Service Guide
Testing Continuity
The continuity test measures the resistance of a tested circuit with 2- wire
method at approximately 0.5 mA and determines whether the circuit is intact.
The beeper emits a continuous beep tone when the input resistance value is
less than the approximate 10 W.
Measurement are displayed as shown below:
Measurement method: 0.83 mA ± 0.2% constant current source, open circuit voltage
limited to < 5 V
Continuity threshold: 10 W fixed
Input protection: 500 V RMS on all ranges
1 Press to toggle to the continuity measurement function.
2 Connect the red and black test leads to the input terminal as shown in
Figure 1- 19.
3 Probe the test points and read the display.
Reading rate Measurement display
Slow 120.000 W range
Medium 400.00 W range
Fast 400.0 W range
CAUTION
Disconnect circuit power and discharge all high-voltage capacitors before testing the
circuit continuity to avoid damaging the multimeter or the device under test.

Table of Contents

Other manuals for Agilent Technologies U3402A

Related product manuals