Identifying the Causes of Error Messages
If your scale displays an error message, use the chart below to
determine their possible cause and solution. If you encounter an
error message and are not able to resolve it, contact Air-Weigh
Support.
ERROR MESSAGE DESCRIPTION POSSIBLE CAUSE
SYSTEM ERROR Deectionsensorvalue
at 0.
Bad/damageddeection
sensor; broken/
disconnected cable.
MEMORY FULL, NO
RECORDING! PUSH
“ESC”
The BinMaxx Comlink
data buffer is full; no new
lift information will be
saved.
Lift data is at maximum
capacity – ERASE ALL to
clear.
CALIBRATION
REQUIRED BEFORE
USE
The BinMaxx scale must
be calibrated.
The scale needs to be
calibrated.
EMPTY CALIB
REQUIRED BEFORE
USE
TheHEAVYcalibration
has been completed, but
not the EMPTY.
The EMPTY calibration
needs to be completed.
BAD CALIB RUN!
PUSH “ESC”
This error can occur
during the calibration lift or
drop (empty or heavy).
The operator took too
much time during the
lift/drop; the lift did not
pass the low/high prox
switch; the user took too
much time before lift/drop
started after the display
ashed;erraticlift/drop.
BAD LIFT! PUSH “ESC”! The scale did not
determine an accurate
weight from the lift. “ESC”
will return you to the last
lift number.
Lift did not go past high
prox sensor; lift did not
go past low prox sensor;
lift was delayed between
high prox and low prox
switches; failed to lower
elevated bin below low
prox switch prior to lift.
NO PRINTER Print Lift command
issued, no printer
detected.
Printer not connected or
not powered.
Troubleshooting | Identifying Error Messages