at least 0.05 to the Output AO1 (Fig. F.4).
2. Ensure the supply voltage is within the range 15…28 V.
3. The measuring device should be of resolution 0.001 V.
4. Set the reference resistance box to 500 ohm.
5. Start ALP software, select menu item Device>Calibration,
in the opened dialogue box select Analog outputs (Fig.
F.2)
6. Click 'Next' to continue and follow the instructions.
7. The output current should be calculated from the
measured output voltage and entered in the displayed
field.