Maintenance and Troubleshooting | 6-19
9. From the Optical Bench dialog box, disable (turn off) Alc On (off =
unchecked). Alc On must be disabled so the analyzer does not make
automatic adjustments while you are setting the Measure and Reference
signals to operate at optimum ratios to each other, as described below.
Click OK then Apply. Do not save the changes to EEPROM.
Figure 6-7.
Optical Bench dialog
box (Model 932S).
Figure 6-6.
Gas Calibration dialog
box (Model 932S).