Maintenance and Troubleshooting | 6-25
If the PMT signals are outside their normal range of operation after the
Auto-Setup is complete, check for the following:
Adjust the Lamp Max setting slightly (typical range is 4500–7800 mV)
and start another Auto-Setup.
Was the Auto-Setup done with Zero gas flowing through the sample
system?
Are the Measuring Cell Windows clean?
Are the PMT signals weak?
Are the PMTs in their original sockets (is the higher gain PMT in the
Measure path)?
Is the PMT Buffer board faulty?
Do the source lamps need replacement (weak lamps)? If not, are the
source lamps aligned?
If the lamps were replaced, is the correct socket connected to the cor-
rect lamp? Lamp 1 is the closest to the Optical Bench board.
Have there been any leaks from the Measuring Cell? If so, check the
condition of the Optical Filters, Beam Splitter, Mirrors, Windows,
Lenses, and o-rings in the Optical Bench and in the Measuring Cell
(including Reflector Block). Contact AMETEK for procedures not dis-
cussed in this manual.
Refer to the “Troubleshooting and Diagnostics” section in this chapter
for alarm conditions and corrective action.