Maintenance and Troubleshooting | 6-39
15. Open the valves isolating the analyzer from the sample stream and/or
remove the cap from the Vent Line outlet.
16. Close the Configurator Software, power down the computer, and
disconnect the serial cable from the analyzer’s Communications Port.
Close the Communications Port cover and hand-tighten the screw to
secure it in place.
17. Using a soft, nonabrasive cloth, gently clean the joining areas (flame-
path) of the Upper Enclosure and its door and replace the (24) M10
screws. Tighten the M10 screws to 9.0 Nm, ±1.0 Nm (80 in.-lb, ±9
in.-lb).
The procedure is complete.