AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 17 / 60
1 Framebus
2 Reference earth connection
3 Signal OUT
4 CH 1 IN
5 CH 2 IN
6 Offset OUT
7 Ventilation
8 Mains Fuse
9 Safety Circuit
10 Mains input
11 Power on switch
12 Mains selector 115V / 230V
8 Mains Fuse
Mains AMP 200N2 AMP 200N1.1 AMP 200N1
115 V 16 AT 16 AT 2 x 8 AT
230 V 10 AT 10 AT 2 x 4 AT
9 Safety Circuit
The Safety Circuit is a 12V loop who switches off the power supply of the amplifier. The ac voltage will be
interrupted by a relays.
Attention: After open the Safety circuit there is still same energy in the storage capacitors
of the amplifier.
10 Mains Input
115VAC/230VAC 16A input connector (build in socket: IEC-60320 C20, compatible plug: IEC-60320 C19).
11 Mains Switch
After switch on the power LED is flashing few seconds, till the AMP200 is ready for use.
12 Mains selector
Selection of 115V / 230V