AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 36 / 60
Test characteristics Default settings as per Standard
Method:
Coupling Method : Radiated
Test Regulation Method: Magnetic field (Calculation)
Diagram Unit X-Axis : kHz ; Log
Y-Axis: dBpT
Tolerances
Tolerances : +10% -0%
Additional Settings:
Additional Level Offset 0V
Level Change at next Testpoint: -3dB
User Software settings for adapt the hardware
The user has to complete the test settings with his individual used coupling and measuring devices. New
coupling devices can be installed in the menu Setup/Coupling Device by the user.
Coupling
Coupling Type Antenna
Coupling Device Radiating Loop
A: Coupling Device:
Selectable 120mm Radiating Loop or a Helmholtz Coil from default list:
Radiating Loop
Radiating Loop (RL 120)
Solar 9230-1
ETS Lindgreen 6703
Schwarzbeck FESP 5132
Helmholtz Coils
Not available
Setup for add or modify new Antennas in the Library in the Setup Coupling Devices
For more information refer to the Autowave software manual.