AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 57 / 60
Magnetic Field Monitor:
Measuring Instrument (AC current)
1. Instrument AMP Internal
2. Instrument value Voltage
Measuring Equipment
3. Equipment Type H-Field sensor
4. Equipment Sensor name
B: Measuring Instrument (Magnetic Field [dBpT]) :
The current measuring is realized with a current probe connected to a measurement receiver. This instrument will
measure the transferred ac voltage. The selected AC current instrument must be able to measure up to 100kHz
Measurement Instrument
Measuring AC voltage
AMP200 Internal Internal
EM Test proposal for loop sensor ( Frequence selective instrument)
Scope
LeCroy preferred manufacturer by EM Test
Keitley 2000
Connect Instrument
Press Connect for check the instrument connection.
Setup : for add or modify an instrument in the library
C: Current Monitor (Antenna current and therefore calculated-Field [dBpT]):
The current monitor is the antenna current measurement with a current probe connected to a measurement
receiver. This instrument will measure the transferred ac voltage. The selected AC current instrument must be
able to measure up to 100kHz.
Note: The most h-Field verification are performed at very low field levels 110 dBpT = 0.25A/m . Using a radiating
loop with an antenna factor of 76.3, the resulting antenna current is in the range 2..3mA. To measure such small
signal it is necessary to use a multimeter or receiverwith a resolution of these levels. A scope may not be
sufficient.
Measurement Instrument
Measuring AC voltage
Keithley 2000 DMM
Voltage measurement (for calculation of the setted level [ dBpT])
Measurement Equipment
Current measurement
Current Probe
Agilent N2783A