AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 57 / 60
Magnetic Field Monitor:
Measuring Instrument (AC current)
1. Instrument AMP Internal
2. Instrument value Voltage
Measuring Equipment
3. Equipment Type H-Field sensor
4. Equipment Sensor name
B: Measuring Instrument (Magnetic Field [dBpT]) :
The current measuring is realized with a current probe connected to a measurement receiver. This instrument will
measure the transferred ac voltage. The selected AC current instrument must be able to measure up to 100kHz
EM Test proposal for loop sensor ( Frequence selective instrument)
LeCroy preferred manufacturer by EM Test
Connect Instrument
Press Connect for check the instrument connection.
Setup : for add or modify an instrument in the library
C: Current Monitor (Antenna current and therefore calculated-Field [dBpT]):
The current monitor is the antenna current measurement with a current probe connected to a measurement
receiver. This instrument will measure the transferred ac voltage. The selected AC current instrument must be
able to measure up to 100kHz.
Note: The most h-Field verification are performed at very low field levels 110 dBpT = 0.25A/m . Using a radiating
loop with an antenna factor of 76.3, the resulting antenna current is in the range 2..3mA. To measure such small
signal it is necessary to use a multimeter or receiverwith a resolution of these levels. A scope may not be
sufficient.
Voltage measurement (for calculation of the setted level [ dBpT])