48 23-03-2009 105446 06
Press
to select the Auto step menu.
F3F2 F4 F5F1
Two functions are available using the soft keys
4.7.1 Running an Auto step calibration
Press
to select Run auto step.
F3F2 F4 F5F1
Press
to enable the editor to change the Auto step setup
and step temperature values.
No of steps: the number of temperature steps per
direction (T
1
T
x
) can be set using integers from 1 – 20.
When a Two-way mode is selected, the same number of
steps are used for the second direction (T
x
T
1
).
Mode: toggle between "One-way" and "Two-way".
Hold time: defines the time (in minutes) the temperature
is maintained (after it is stable) for each step.
T step values: must be set within the sensors permitted
range.