70   23-03-2009  105446  06 
7.0 Maintenance 
7.1 Cleaning 
 
Caution… 
Before cleaning the calibrator, you must switch it off, allow 
it to cool down and remove all cables. 
 
Users should/must carry out the following cleaning procedures as and 
when required: 
 
 The exterior of the instrument - Clean using water or isopropyl 
alcohol and a soft cloth.  
The cloth should be wrung out hard to avoid any water penetrating 
the calibrator and causing damage.  
The keyboard may be cleaned using isopropyl alcohol when 
heavily soiled. 
 
 The insertion tube - must always be clean and should be 
regularly wiped using a soft, lint-free, dry cloth.  
You must ensure there are no textile fibres on the insertion tube 
when it is inserted in the well. The fibres may adhere to the well 
and damage it.  
If the calibrator has reached a temperature below 0°C/32°F, ice 
crystals may form on the insertion tube. This, in turn, may cause 
verdigris to form on the material. 
To prevent this from happening, the insertion tube must be dried. 
This is done by heating up the calibrator to min. 100°C/212°F and 
any water left will evaporate. 
Remove the insulation plug while heating up. 
It is very important that humidity in the insertion tube is removed to 
prevent corrosion and frost expansion damages. 
 
 The well - must always be clean.  
Dust and textile fibres in the well should be removed from the dry-
block calibrator using e.g. compressed air.  
Remains of silicone oil in the well should be removed from the 
liquid bath calibrator using a special cleaning oil.